Setting cleaning validation acceptance limits and developing an HPLC method for determination of risperidone residues on manufacturing equipment

Milena Prculovska1✉, Katerina Kochova1, Irena Slaveska Spirevska1, Angela Nikolovska1, Monika Jakovleska Ruseska1, Elena Petrovska Kostadinoska1

  1. 1 Replek Farm, Skopje, Str. Kozle 188 Skopje, Republic of North Macedonia

Received 2 November 2026 · Accepted 2 November 2026 · Published 2 November 2026

DOI: https://doi.org/10.5281/zenodo.23887565

Article PDF · Original poster (PDF)

views
9
downloads
4
countries
5

Since publication · journal site and Zenodo archive · robots excluded

Key messages

  1. Risperidone residues require stringent cleaning validation to prevent cross-contamination.
  2. A sensitive RP-HPLC method was developed and validated for risperidone residue detection on equipment surfaces.
  3. PDE-based acceptance limits provide a practical approach for cleaning validation and risk control.

Abstract

Introduction

Risperidone is a potent atypical antipsychotic; residual carryover therefore requires stringent control. Cleaning validation protects product quality by preventing cross-contamination between manufacturing campaigns.

Materials and methods

Sampling was performed using Textwipe 714A swabs with direct sampling of manufacturing-equipment surfaces. Risperidone certified reference material (CRM) was used. The HPLC system was Shimadzu Nexera XR UPLC with a LiChrospher RP Select B column (125 × 4.0 mm, 5 µm). The mobile phase was ACN / 0.25% KH2PO4 = 34:66 (v/v) at pH 3.0. Elution was isocratic at 1.0 mL/min and 35 °C. Detection was at 235 nm with injection volume 50 µL. Diluent was methanol / 0.08% H3PO4 = 7:3 (v/v). Validation design included selectivity, accuracy, linearity, precision, and quantification limit.

Results

The method met the reported sensitivity, linearity, and precision criteria for routine residue monitoring. Linearity showed R² = 0.9990 over seven concentration levels ranging 0.12–3.6 µg/mL. Limit of quantification was 0.12 µg/mL, precision RSD < 5% at LOQ. Selectivity was demonstrated by chromatographic traces distinguishing risperidone response from matrix and solvent contributions. Acceptance limits were established using the Permitted Daily Exposure (PDE) approach to control cross-contamination risk.

Conclusion

The validated RP-HPLC method is suitable for monitoring risperidone in equipment-surface swab samples. PDE-based limits were described as realistic and attainable, supporting routine cleaning validation and cross-contamination control.

Original poster

References

  1. ICH Q2(R1) (2005), Validation of Analytical Procedures; LeBlanc (2000), Validated Cleaning Technologies; Schotte et al. (1993), Brain Research 631, 191-202.
Keywords
Risperidone · Cleaning validation · HPLC · Residue monitoring · PDE
Received
2 November 2026
Accepted
2 November 2026
Published
2 November 2026
Peer review
Single-anonymized review by the Editorial Board · How we review
DOI
https://doi.org/10.5281/zenodo.23887565
License
© The Authors 2026. Distributed under CC BY 4.0
Journal
MP Journal of Medical Research · ISSN 3144-8402

Related posters